Lifeline PRO AED - Semi-automatic Defibrillator

SKU: 102364
£2,019.60
£2,423.52 (inc VAT)
Delivery Timeframes: 3-5 WORKING DAYS

Our Lifeline PRO AED - Semi-automatic Defibrillator.

Quantity:
1
Out Of Stock

Out of stock

TYPE
Semi-Automatic external defibrillator 
with Manual Override
MODEL
 DDU-2400 series
WAVEFORM
Biphasic Truncated Exponential
(Impedance compensated)
ENERGY
AED Mode:
  Adult: 150 Joules
  Child / Infant: 50 Joules
  (Nominal into 50 Ohm load)
Manual Mode:
  User selectable from 25-200 Joules
  (Nominal into 50 Ohm load)
Child/Infant:
  User selectable from 25-100 Joules
  (50 Joules only when using DDP-2002
 attenuating defibrillation pads)
CONTROLS
Lighted On/Off button
Lighted Shock button
Menu Selection buttons
CHARGE TIME
AED Mode:
  4 seconds or less
  (from shock advised)**
Manual Mode:
  9 seconds or less (150 Joules)***
  12 seconds or less (200 Joules)***
DISPLAY
High-resolution color LCD
VIDEO PROMPTS
Full motion video
On-screen text prompts
CPR COACHING
Video and voice coaching
On-demand video help
VOICE PROMPTS
Extensive voice prompts guide user through operation of the unit
RESCUE PROTOCOL
AHA/ERC 2010
Supports protocol updates by the user
(password protected)
  ** Typical, with new battery at 25° C
*** Typical, new battery depleted by 6 shocks at 25° C
Self Tests
AUTOMATIC
Automatic daily, weekly and monthly circuitry tests
BATTERY INSERTION
System integrity test on battery insertion
PAD PRESENCE
Pads preconnected tested daily
USER-INITIATED
Unit and battery pack system test initiated by the user
STATUS INDICATION
Visual and audible indication of unit status
STATUS SCREEN
Unit self-test results
Pads and battery information
(status and expiration)
Patient Analysis System
PATIENT ANALYSIS
Automatically evaluates patient impedance for proper pad contact. Monitors signal quality and analyzes patient ECG for shockable/non-shockable rhythms.
SENSITIVITY/SPECIFICITY
Meets AAMI-DF-80 specifications and AHA recommendations
Defibrillation / Monitoring Pads
MODEL
Adult – DDP-2001
Child / Infant – DDP-2002
TYPE
Pre-connected, single-use, non-polarized, disposable, self-adhesive electrodes with cable and connector
SURFACE AREA
77 cm2 (nominal, each pad)
50 cm2 (nominal, each pad)
Environmental
TEMPERATURE
Operating: 0 to 50°C (32 to 122°F)
Standby: 0 to 50°C (32 to 122°F)
RELATIVE HUMIDITY
Operating / Standby: 5%-95%
(non-condensing)
ALTITUDE
-500 to 15,000 ft (-150 to 4500m) per MIL-STD-810F 500.4 Procedure ll
VIBRATION
Ground (MIL-STD-810F 514.5 Category 20)
Helicopter (RTCA/DO-160D, Section 8.8.2, Cat R. Zone 2, Curve G)
Jet Aircraft (RTCA/DO-160D Section 8. Cat H, Zone 2, Curves B & R)
SHOCK / DROP ABUSE TOLERANCE
MIL-STD-810F 516.5 Procedure IV 48 inches (1.2 meters), any edge, corner, or surface, in standby mode
CRUSH TEST
1,000 pounds (450 kg)
SEALING / WATER RESISTANCE
IEC60529 class IP55; Protected against dust and waterjets (battery pack installed)
ESD
EN61000-4-2: (Open air up to 8kV or direct contact up to 6kV)
EMC (Emission)
CISPR 11 Group 1 Level B
and FCC Part 15
EMC (Immunity)
IEC 61000-4-3 and
IEC 61000-4-8

Operating Modes

AED MODE: With Video Display (animated instructions and CPR coaching) or with ECG Display.
MANUAL MODE: Displays high-resolution ECG data and event information and defibrillation energy selection from 25-200 Joules with charge, shock and disarm control.
ECG MONITOR MODE: Displays high-resolution ECG data and event information (with 3-lead electrodes).
Battery Pack
MODEL
DBP-2003, DBP-2013 (aviation)
POWER
12V, 2800 mAh
TYPE
Lithium/Manganese Dioxide Disposable, recyclable, non-rechargeable
CAPACITY **
125 shocks or 8 hours continuous operation
STANDBY LIFE **
4 years
LOW BATTERY INDICATORS
Visible
Audible
** Typical, with new battery at 25° C
INTERNAL EVENT RECORD
Critical ECG segments and rescue event parameters are recorded (greater than 60 minutes) and can be downloaded to a removable data card
PC-BASED EVENT REVIEW
ECG with event tag display, and audio playback when available
REMOVABLE STORAGE
(optional) Up to 30 hours of ECG and event data storage (no audio option) or up to 3 hours of audio (audio option). ECG and event storage on a removable data card. Actual length of storage is dependent on card capacity.
USB PORT
Event download and maintenance operations

Delivery costs for orders are calculated by the price, weight and volume of the item and this will be displayed to you within the shopping basket as you add items to the basket.